Description
Looking to buy a multi-panel drug test in Chisinau or anywhere in Moldova?
This testing panel is designed for rapid qualitative detection of the following drugs and their metabolites in human urine: amphetamine (AMP), marijuana (THC), morphine (MOP), methamphetamine (MET), barbiturates (BAR), benzodiazepines (BZO), cocaine (COC), phencyclidine (PCP), methadone (MTD), and ecstasy (MDMA).
This device is intended exclusively for urine testing. Its performance with any other type of specimen is not validated. Adulterated or contaminated urine samples may result in incorrect outcomes. Strong oxidizing agents (such as bleach) may interfere with test results. If sample tampering is suspected, collect a fresh specimen.
The test provides qualitative screening only. It does not measure drug concentration or intoxication level. Technical errors, interfering substances, or medications may lead to false results. A negative result does not always mean the absence of drugs — concentrations may simply be below the cut-off level. Certain medications or supplements may cause false positives.
COLLECTION AND PREPARATION OF THE SAMPLE
Collect urine in a clean container. Store the sample refrigerated (2–8°C) for up to 48 hours or freeze (−20°C or lower) for longer storage. Before testing, allow the sample to reach room temperature.
Use only clear, unobstructed urine samples.
TEST PROCEDURE
Perform the test at room temperature (18–30°C).
Open the sealed pouch at the notch.
Remove the test panel.
Immerse the absorbent tip into the urine for approx. 10 seconds. Do not exceed the MAX line.
Lay the test on a clean, dry, flat surface.
Read the result at 5 minutes.
Do not interpret results after 10 minutes.
RESULT INTERPRETATION
Positive (+)
A control line appears, while no test line appears — the corresponding drug is detected.
Negative (–)
Both the control line and the test line appear — the substance is absent or below the detection limit.
Invalid
If no control line appears, the test is invalid and must be repeated with a new device. If issues persist, contact the distributor or retailer with the batch number.
Note: Line color intensity does not affect interpretation.